Automated working distance adjustment for a handheld OCT-Laryngoscope
SPIE Photonics West: BiOS
1.-6. Februar
San Francisco
2014
Type: Konferenzbeitrag
DOI: 10.1117/12.2035790
Abstract
Optical coherence tomography (OCT) is an imaging technique which enables diagnosis of vocal cord tissue structure by non-contact optical biopsies rather than invasive tissue biopsies. For diagnosis on awake patients OCT was adapted to a rigid indirect laryngoscope. The working distance must match the probe-sample distance, which varies from patient to patient. Therefore the endoscopic OCT sample arm has a variable working distance of 40 mm to 80 mm. The current axial position is identified by automated working distance adjustments based on image processing. The OCT reference plane and the focal plane of the sample arm are moved according to position errors. Repeated position adjustment during the whole diagnostic procedure keeps the tissue sample at the optimal axial position. The auto focus identifies and adjusts the working distance within the range of 50 mm within a maximum time of 2.7 s. Continuous image stabilisation reduces axial sample movement within the sampling depth for handheld OCT scanning. Rapid autofocus reduces the duration of the diagnostic procedure and axial position stabilisation eases the use of the OCT laryngoscope. Therefore this work is an important step towards the integration of OCT into indirect laryngoscopes. © (2014) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.